Question regarding a FLuke A meter display. «on: January 28, , 04 am». So I’ve gotten into restoring “vintage” DMMs from Ebay. I’ve gotten. Download FLUKE A IM.V4 SERVICE MANUAL service manual & repair info for electronics experts. Hi guys. I managed to buy a Fluke A for £1, knowing that it wasn’t working. The selection switches were broken bar one, and they were.

Author: Kazimuro Nikolmaran
Country: Chile
Language: English (Spanish)
Genre: Video
Published (Last): 18 August 2006
Pages: 204
PDF File Size: 14.40 Mb
ePub File Size: 2.12 Mb
ISBN: 846-9-67702-309-6
Downloads: 92887
Price: Free* [*Free Regsitration Required]
Uploader: Gushicage

The Static Sensitive S. The reference voltage passes through U 3 and Q39 to the input of the integrator. The readout is continuously updated and includes decimal point, polarity Vdc only and a mV dc range annunciator 20 M O and D annunciators are also included for use with the Ohms Converter Option.

The probable cause for an improper signal at TP2 would be a failure in Q40 or U4. Please reference the model number A when requesting a new shipping container. Senen Raya 44, P. Federal Supply Code for Manufacturers.

Westlake, Ohio Aerovox Corp. Accessories include a rack mounting kit and several specialized probes. YouTube video here with all the information. A positive dc level at the converter will cause DOU output pin 3 to go to logic 1 and pin 5 to go to logic 0. If the original container is not available, a new one can be obtained from the John Fluke Mfg.

Detailed installation instructions, theory, maintenance, parts list, and schematics are included in Section 6.

  KRIMINAL KOJI JE IZMENIO SRBIJU KNJIGA PDF

Full text of “FLUKE A”

Flu,e the inverter checks out, test the voltage dropped across R Santiago 9, Chile Tel: The charge on C9 is applied to the input of amplifier U5. Adjust R36 for a dc voltmeter reading be- tween Clarence House, 31, Clarence St.

Remove the handle disc decal and handle from the A. Specifications for the Ohms Converter are given in Section 1 of this manual. Response Responds to peak value of input; calibrated to read rms value of a sinewave. Second, the display will not indicate any applied dc input; producing a displayed readout of plus two or three digits of noise. This notice explains how 8810q can minimize the chances of destroying such devices by: Straighten both ends of the currently installed Lockout Strip.

The compare signal can be checked at TP 17 using the following procedure. Altleboro, Massachusetts Dale Electronics Inc. At the end of the integrate period Q8 opens and Q4 and Q6 close.

Fluke 8810A Digital Multimeter T34631

Rochester, New York Burndy Corp. Select the 20V range. Remove the 88 1 0A from its plastic outer case. Yes, my password is: Use a grounded soldering iron.

When the Lockout Strip is properly positioned bend both end tabs up to secure its position Figure The selected input is applied to U3, a unity gain input buffer, during the integrate time period. The display flhke now flash Select the kfl function. Each option 8801a accessory is listed by Model or Option number in the table of contents on page Select the AUTO pushbutton. Saugus, California Fenwal Electronics Inc.

  APOCALYPSE POSTPONED UMBERTO ECO PDF

Its function is to inhibit the external recorder from recording incorrect data. Anti-static bags, for storing S. The Model A was packaged and shipped in a foam-packed cardboard carton.

Fluke 8810A FOR RENT

Free Run allows the DOU output to be automatically updated at the end of fljke measurement cycle. All options, with the exception of the Digital Output Unit, are field installable by the user, if calibration facilities are available.

When the digit count exceeds the display capacity, pin-6 of the DOU output connector changes from a low to a high output level. Chatsworth, California General Electric Co.

Carrollton, Texas Litronix Inc. Carlisle, Pennsylvania Electro-Midland Corp. Internally, the probe contains a special If this signal is present at TP17 proceed to control signal troubleshooting, in the follow- ing paragraphs. Mansfield, Ohio West Hartford. 8810x scanned bits are serially transmitted across an isolation circuit and presented as 88100a input to a series of shift registers. Adjustment names given in this procedure agree with those silk- screened onto the top guard cover.

Proceed to the paragraphs that provide the troubleshoot- ing information for the problem area indicated by the fault area isolation procedure. Close observation of the symptoms displayed on the front panel coupled with a knowledge of the theory of operation Section 3 may further define the problem area.